6
Troubleshooting
A system fault will be indicated by an acoustic signal and one of the follow-
ing fault codes on the main display:
Display Fault
E03 E
Overheated coil
E05 E
Temperature increase of
less than 1° every 30
seconds (or 1° every 60
seconds)
E06 E
Temperature probe(s) not
connected (or defective)
E07 E
Failure during current
measurement
E13 E
Temperature probe pulled
out in Delta-T mode
7
Spare parts
Art. No.
Description
190-12240
Yoke 42,5 x 42,5 x 219mm
(for bearings with minimum 60mm bore)
190-12230
Yoke 28 x 28 x 219mm
(for bearings with minimum 40mm bore)
190-12220
Yoke 14 x 14 x 219mm
(for bearings with minimum 20mm bore)
190-12260
Support set for IH 045
190-10020
Protective gloves
190-10010
Temperature probe Typ K
40
Action
Wait until the inductive coil cools.
Check the temperature probe connec-
tion. If the connection is OK, select
the 60 second interval as described
in section 5 or operate the heater in
TIME MODE.
Check the temperature probe(s).
Return IH 045 for repair.
Check the temperature probes.
Induction Heater IH 045
© simatec ag