14. Fill the storage tube to the halfway point with pH 4 buffer solution.
15. Submerge the pH sensor in the pH 4 buffer solution inside the
storage tube. Start the pH4 calibration in the iAquaLink app or on the
OneTouch controller (AquaLink
SYSTEM SETUP > TRUSENSE, see Figure 9.
Figure 9.
Submerge sensor & start calibration
16. Remove the pH sensor from storage tube and rinse the tip of the sensor
with clean water.
17. Once calibration is complete, carefully insert pH sensor back into
Jandy TruSense and hand tighten the compression fitting.
NOTE: DO NOT leave sensor exposed to air.
It may take a few hours for the pH & ORP Sensor readings to
stabilize after being installed on the pool plumbing.
5.2
1 point pH Sensor Calibration
1.
Get a clean glass or plastic container and fill it with water. DO NOT
use a metal container.
2.
Measure the pH of water sample.
3.
Input measured pH in the iAquaLink app or on the OneTouch
controller (AquaLink
SETUP > pH > 1 POINT.
Jandy
TruSense Water Chemistry Analyzer | Owner's Manual
®
RS only) go to: MENU / HELP >
®
RS only) go to: MENU / HELP > SYSTEM
®
Calibrate pH