Control output accuracy:
Remote Off:
DIGITAL INTERFACES (XPF 60-20P only)
Full digital remote control facilities are available through the RS232, USB, LAN and GPIB
interfaces.
Voltage Setting:
Current Setting:
RS232:
GPIB:
USB:
LAN:
Remote Command
Processing Time:
GENERAL
AC Input:
Power Consumption:
Operating Range:
Storage Range:
Environmental:
Safety:
EMC:
Size:
Weight:
6
Voltage: 0.3% ±20mV
Current: 0.5% ±50mA.
Output impedance: 125Ω.
Rear panel connections allow a switch closure to turn output off. Remote Off
is referenced to the output control circuit and will therefore be within ~1V of
the positive output potential.
16-bit, Resolution 10mV, Accuracy ± (0·05% +10mV)
16-bit, Resolution 1mA, Accuracy ± (0·3% + 0·005A)
Standard 9-pin D-connector. Baud rate 9600.
Conforming with IEEE488.1 and IEEE488.2
Standard USB 2.0 hardware connection. Operates as a virtual COM port.
Ethernet 100/10base-T hardware connection. LXI V1.2, Class C compliant.
Typically <25ms between receiving the command terminator for a step
voltage change at the instrument and the output voltage beginning to change.
110V – 240V AC ± 10%, 50/60Hz.
Installation Category II.
600VA max.
+5ºC to +40ºC, 20% to 80% RH.
−40ºC to + 70ºC.
Indoor use at altitudes up to 2000m, Pollution Degree 2.
Complies with EN61010-1.
Complies with EN61326.
107mm x 130mm x 400mm (WxHxD) ¼-rack width x 3U height
(optional rack mounting kit available).
4.25kg