Ametek Sorensen XPF Serie Manual De Instrucciones página 12

Ocultar thumbs Ver también para Sorensen XPF Serie:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

USB (XPF 60-20P only)
The USB port is connected to instrument ground. It conforms with USB 2.0 (Full Speed) and
accepts a standard USB cable. The Windows plug-and-play functions should automatically
recognise that the instrument has been connected. If the correct driver is not found, follow the
Windows on-screen prompts and install the required files from the CD supplied.
LAN (XPF 60-20P only)
The LAN interface is designed to meet LXI ( Lan eXtensions for Instrumentation) version 1.2; the
instrument is Class C compliant. Remote control using the LAN interface is possible using a
TCP/IP Socket protocol. The instrument also contains a basic Web server which provides
information on the unit and allows it to be configured. Since it is possible to misconfigure the LAN
interface, making it impossible to communicate with the instrument over LAN, a LAN Configuration
Initialise (LCI) mechanism is provided via a recessed switch on the rear panel (marked LAN
RESET) to reset the unit to the factory default.
Further details are given in the Remote Operation chapter. For more information on LXI standards
refer to www.lxistandard.org/home
GPIB (XPF 60-20P only)
The GPIB signal grounds are connected to the instrument ground. The implemented subsets are:
The GPIB address is set from the front panel.
SH1 AH1 T6 TE0 L4 LE0 SR1 RL2 PP1 DC1 DT0 C0 E2
11
Tabla de contenido
loading

Este manual también es adecuado para:

Sorensen xpf 60-20Sorensen xpf 60-20aSorensen xpf 60-20p

Tabla de contenido