5. INSTALLATION
5.1 SETTING UP THE LIGHT SOURCE
Place the light source on a stable surface (cart, counter, stand, etc.).
NOTE
Avoid places where the light source may be splashed with liquid.
Absolutely DO NOT use in any environment with explosive or flammable
gases.
DO NOT block the venting grids of the light source or impede exhaust air
flow. Users must ensure that ventilation air temperatures are below the
maximum rating for the light source and are not influenced by warm
exhaust from other sources.
Make sure the power switch is in OFF position.
Connect AC power cord to the power inlet located on the rear panel of the light source.
Make sure power cord is completely seated into the power inlet.
Use only cords provided with the light source.
CAUTION
Plug the AC power cord into a wall outlet using the three (3) prong plug supplied with the unit.
To prevent electric shock, connect power cords of peripheral equipment
through medical isolation transformers.
CAUTION
NOTE
When using medical isolation transformer, be sure to check the
transformer power ratings.
Make sure that the power cord is connected to the main power with three
prong plug (USA use UL2601-1 rated isolation transformers and/or power
strips only).
When used in clinical or residential areas near radio or TV units,
this equipment may be subjected to radio interference.To avoid
adverse electromagnetic effects, DO NOT operate this equipment
CAUTION
near RF energy equipment.
5.2 CONNECTING THE LIGHT CABLE
Connect the light cable to endoscope and then plug the light cable endfitting into the
appropriate port of the turret on the front panel.
®