Sky CI M90 sound processor does not respond to the
implant device because of an unusual field disturbance,
move away from the disturbing field. Consult Guidance
and Manufacturer's Declaration section in this IFU for
determining safe EM field levels.
•
Use of the Sky CI M90 sound processor next to or
stacked with other equipment should be avoided
because it could result in improper operation (e.g.,
poor sound quality, cutouts in sound, delay in
response when making adjustments, or loss of lock). If
such use is necessary, the Sky CI M90 and the other
equipment should be observed to verify that they are
operating normally.
Use of components, accessories, and cables other
•
than those specified in the Compatibility section of this
document could result in increased electromagnetic
emissions or decreased electromagnetic immunity
of this equipment and result in improper operation
(e.g., poor sound quality, cutouts in sound, delay in
response when making adjustments, or loss of lock).
Portable RF communications equipment (including
•
devices such as antenna cables and external
antennas) should be used no closer than 12 inches
(30 cm) to any part of the Sky CI M90 sound
processor, including the headpiece cable. Decreased
performance of the sound processor, headpiece,
or other accessories could result (e.g., poor sound
quality, cutouts in sound, delay in response when
making adjustments, or loss of lock).
6