3) As Fig.6, when the therapy is finished, take off the tube. Press down the external fixation
steel rim, and take out the tube joint.
Steel rim
EN
Note: Nebulizer must be cleaned after each use. Clean nebulizer and connection tube
a soft cloth dampened with detergent and clean water; for the nebulization nozzle and
mask use clean water to clean first, then disinfecton and sterilize by dipping into medical
alcohol for 5min, and clean water, and finally put them in the packet after drying. (Detailed
clean method please read the simple nebulizer user manual)
6. MAINTENANCE
ROUTINE MAINTENANCE
1. While powered off clean the outside case with soft damp towel and little detergent,
and then wipe it up with dry towel, once or twice per month.
2. Air intake filter is used to prevent dust, and need to be cleaned regularly. It is a critical
maintenance step to clean the filter, at least once a week depending on environmen-
tal conditions. Operation steps as below.
1) Take off the two filter nets on both sides of the case, take out the sponge mesh.
2) Rinse the sponge mesh with clean water. Get ride of extra water and dry naturally.
3) Set back after drying for future use.
4) Re-install the sponge the and the filter nets.
NOTE:
• If replacement is needed, please use accessories supplied by Sysmed. Accessories de-
signed by other manufacturers have not been tested by Sysmed and are not recom-
mended for use with Sysmed products.
• Do not operate the unit without the air intake filter in place.
3. Over-load protector
• Over-load is caused by over heating, after checking and
trouble shooting, press "reset" to re-start the concentrator.
4. Cleaning the Humidifier and cannula
1) Clean the cannula every day. First clean with detergent and then rinse with clean water
completely, and dry naturally. Changing a new cannula once a month is recommended.
2) The distilled water or cold boiled water in humidifier should be refreshed every day.
The humidifier should be cleaned and disinfected every 3 weeks, clean the humidifier
with detergent and then rinse with warm or hot water, dip it in household disinfectant for
20-30 minutes, clean again with warm or hot water, and then dry naturally. If not use, put
it in a clean bag.
3) If oxygen tube or cannula is needed, please contact your local home care provider or
dealer for recommendations. They should also give you instruction on the proper usage,
maintenance, and cleaning.