No.
Type of
Causes of error
error
Open
frontal
Frontal contact is damaged
Alarm
8
"
WORK
POSITION
"
(see Chapter
Main control is damaged
4.5 pto. D)
Light not working properly
Light does
9
not turn on
Led bar is not working
Glass not in closed position
UV
germicidal
lamp does
10
UV lamp does not work
not turn on
(optional
accessory)
Reactance is damaged
Front position security
Intermittent
micro activated
11
alarm
activated
Main control is damaged
Power cord cable
disconnected
Display
12
does not
No power
turn on
Main control is damaged
WARNING
All repair and maintenance work in the cabinet should be done by qualified technical staff
with knowledge about the maintenance procedures required by the cabinet.
UM-BioIIAdvancePlus-EN-1812
Correction of error
Slide down the glass to closed position and
check that the frontal contacts are not
loose and remove any type of obstruction.
Notify the Technical Service to receive
instructions for adjusting the contacts.
Call a specialised technician to replace the
control.
Check that the connection in the control
card is correct (Annex A7).
Check that the lamp is properly connected.
Change the led bar.
Check that the front slides properly.
Check that the lamp holder is connected
properly.
Check that the connection in the control
card is correct (Annex A7).
Change the UV lamp.
Change the reactance.
Open the front and check the security
micro is not loose.
Call the Technical Service to receive
instructions on adjusting micros.
Call a specialised technician to replace the
control.
Check that the power cord cable is
properly connected in cabinet and in
socket on the wall.
Check voltage on the socket of the wall.
Call a specialised technician to replace the
control.