There is a time delay before the unit performs the
"Manual Scan". This time delay allows the light source
to be turned on before "Manual Scan" starts. Time
delay can be set in the SETUP menu.
Single Point Test
1. Press
button to turn on the Analyzer.
2. Properly connect the pair of Kelvin clips to the solar panel and the Analyzer. The red Kelvin
clip is for positive pole and the black Kelvin clip for negative pole.
3. Press
button to enter the SETUP menu: (Type in the testing current)
Point:609mA
After setting up the current value, press
button to start Testing a Single Test. After the testing is finished, the result
4. Press
will come out like below. The result (P, V, I) is shown in reverse video as below.
Time delay in the "Single Point Test" allows the current
simulation to last longer. Though the max. value is
9.999 sec., the time delay is changed to 10 msec. if
power is over 100 W. The time delay is extended to 3
seconds if power is less than 100 mW.
Result of Single Point Test
©Copyright Abacanto Digital SA.
button again to exit the SETUP menu.