PROBLEM
The display panel does
not illuminate when the
unit is switched on.
No liquid flowing to the
handpiece when the
footswitch is activated.
No water flow to insert tip
No/poor ultrasonic
vibration
TROUBLESHOOTING
CAUSE
The power cord is disconnected.
The internal fuse has burned out.
Water flow is set too low.
Connections are not secure.
Water filter is blocked.
Water adjustment knob is not
properly set.
Insert is clogged.
Dental office water supply valves
are closed.
Waterline filter is clogged.
The insert is not seated securely
in the handpiece.
Incorrect kHz insert.
The insert is worn.
Power has not been properly
adjusted for the particular insert.
The foot control is disconnected.
Handpiece failure.
Foot control failure.
SOLUTION
Check all the connections of the control unit.
Check the power supply to the dental office.
Contact your dealer or Hu-Friedy.
Adjust the liquid flow water knob.
Check the connection between the handpiece cord
and control unit.
Check the connection between the water tube and
the control unit.
Check the connection between the foot control and
the control unit.
Check the water filter for blockage.
Properly set water adjustment knob.
Contact your dealer or Hu-Friedy.
Open dental office water supply valves.
Replace waterline filter.
Check the insert to ensure proper insertion
into handpiece.
Replace with proper kHz insert.
Replace the insert.
Adjust the power setting to the recommended power
range indicated on the package insert.
Connect the foot control correctly.
Contact your dealer or Hu-Friedy.
Contact your dealer or Hu-Friedy.
SWERV
3