vIDEO-DIAGNOSTIC fuNCTION
The Maia lamp has a function that enables it to be used when filming with a television camera
and/or using diagnostic instruments (Diagnodent and laser, for example) without causing
interference that could alter the diagnostic result.
This function is only in manual switch equipped versions.
Activation of the Video-Diagnostic function:
1. Switch on the Maia dental lamp (a beep will be heard when the control is used).
2. Release the control.
3. Use the control again to reach the minimum light intensity (a beep will be heard when
minimum intensity is reached) then without releasing the control keep it active for at least
4 seconds.
4. A beep is emitted as confirmation, the light intensity rises to the maximum level and the
Video-Diagnostic function is ACTIVE.
If the lamp does not react as described in point 4 above, repeat the whole procedure from point 1.
Deactivation of the Video-Diagnostic function:
1. Switch on the Maia dental lamp (a beep will be heard when the control is used).
2. Release the control.
3. Use the control again to reach the minimum light intensity (a beep will be heard when
minimum intensity is reached) then without releasing the control keep it active for at least
4 seconds.
4. A beep is emitted as confirmation, the light intensity rises to the maximum level and the
Video-Diagnostic function is DEACTIVATED.
If the lamp does not react as described in point 4 above, repeat the whole procedure from point 1.
Dimming of the light intensity with the Video-Diagnostic function ACTIVATED:
With the Video-Diagnostic function activated, the regulation of the light intensity is modified
from a continuous variation to a stepwise variation.
Two intermediate levels of light intensity can be chosen between the maximum and minimum.
Procedure:
1. Switch on the Maia dental lamp (a beep will be heard when the control is used)
2. Release the control.
3. Use the control again to reduce the light intensity and release the control at the intensity
desired.
Note:
• On reaching minimum intensity, a beep will be heard.
• When the dental lamp is switched on again it will return to the maximum light intensity (a
beep will be heard when the control is used).
21