Operating Mode Measure "Meas - WIKA Mensor CPC2000 Manual De Instrucciones

Ocultar thumbs Ver también para Mensor CPC2000:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

6. Commissioning, operation
Function
Leak test
6.6

Operating mode Measure "MEAS"

The Measure mode is used to measure a differential or gauge pressure. The pump is
not active in this operating mode. The pressure is measured directly from the internal
reference sensor.
Setting
1. Press the [MENU] button.
2. Confirm the menu item MODE with [SELECT] and press the [▲] / [▼] buttons until
MEAS is displayed.
3. Confirm with [SELECT].
For the MEAS mode, the submenu items RANGE and STEPS are not relevant.
4. In the menu, select the item UNIT with the [▲] or [▼] buttons.
5. Select the pressure unit.
6. Confirm with [SELECT].
7. With the[MENU] button, exit the menu.
SETTINGS submenu
1. With the [▲] or [▼] buttons, select the menu item SETTINGS.
2. Confirm with [SELECT].
3. In the submenu, select the item ZERO.
4. Confirm with [SELECT].
Under the point, AUTO-ZERO, the automatic zero point correction for the MEAS
mode can be activated or deactivated.
WIKA operating instructions, model CPC2000
Meaning
By pressing the [LEAK] button, the control process will be deactivated
and the system sealed. It can be controlled via the display if the value
is stable. By pressing the [LEAK] button again, the mode will be
stopped and the instrument placed back in Control mode.
EN
21
Tabla de contenido
loading

Tabla de contenido