6. Commissioning, operation
6. Commissioning, operation
Personnel: Skilled personnel
EN
Only use original parts (see chapter 12 "Accessories").
6.1
Location
An unsafe stand or vibration can influence the measurement or lead to injury.
Make sure that the low-pressure controller is placed on a secure surface during
■
operation.
Make sure that the surface is stable.
■
6.2
Mechanical mounting
CAUTION!
Damage to the instrument
To avoid possible damage to the CPC2000 or to the test assembly, pay
attention to the following:
▶
Only connect test and calibration installations once the system has been
depressurised (atmospheric).
▶
The maximum pressure is P
▶
This pressure must NEVER be exceeded.
6.3
Measurement principle
Microprocessor
Actual pressure
value
Pressure value
RS-232 PC interface
USB interface
Display
18
Set-actual value
comparison
Controllers
Pressure
measurement
= 1 bar atmospheric pressure
max
-
Pump
+
WIKA operating instructions, model CPC2000
Valves
neg.
pressure
pos.
pressure