Varex Imaging XRpad2 HWC-M Digital X-ray Manual De Usuario página 65

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 181
Inspection and Maintenance
19.1
Daily Inspection
Perform the following inspection daily. If there is any problem, immediately ask your
establishment's safety representative to contact your dealer, distributor, or device manufacturer.
19.1.1
Before Turning On the Power
Inspection
Check all cables, if
applicable, (Power and
communication tethered
cord, DC-cable, Ethernet
cable, Sync cable) to ensure
that they are not damaged
and the insulation is not
damaged.
Check all connector plugs
and locks to ensure they are
not loose.
Check that the detector is
not damaged.
Check the manufacturing
date of the battery pack to
ensure that its age is not five
years or older.
Check that the battery pack
(wireless mode) or the
Protective Insert (wired
mode) is in the battery
compartment and the
sliding latches are closed.
Check that the battery pack
is not damaged.
Check that the detector is
not loose and all screws are
fixed.
62
Result
Date
Date
/
/
Good/Bad Good/Bad Good/Bad Contact your dealer,
Good/Bad Good/Bad Good/Bad Fully insert the cables and
Good/Bad Good/Bad Good/Bad Contact your dealer,
Good/Bad Good/Bad Good/Bad Replace the battery pack with
Good/Bad Good/Bad Good/Bad Slide the battery pack or the
Good/Bad Good/Bad Good/Bad Replace the battery pack with
Good/Bad Good/Bad Good/Bad Contact your dealer,
www.vareximaging.com
XRpad2 HWC-M Detectors User Manual
Remedy
Date
/
distributor, or device
manufacturer if there is a
problem.
lock them.
distributor, or device
manufacturer if there is a
problem.
a new one and recycle the old
one.
Protective Insert into the
battery compartment as
described in
"Section 13.4,
Removing the Battery Pack"
on page
47.
a new one.
distributor, or device
manufacturer if there is a
problem.
loading