Varex Imaging XRpad2 HWC-M Digital X-ray Manual De Usuario página 38

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 181
XRpad2 HWC-M Detectors User Manual
Table 9
X-ray Detector Specification (Continued)
XRpad2 X-ray Detectors
Image Transfer Time
On-board Memory
Mechanical
Size - (ISO 4090)
Active Area - (mm)
External Dimensions - (mm)
Weight - kg (lbs.)
Housing
Communication
Status Display
Wireless Data I/F
Wired Data I/F
X-ray I/F
Image Performance
Limiting Resolution
Typical MTF
Typical DQE
Typical Radiographic Lag
a. Country-dependent.
b. per IEC 62220-1
3025
Wired: 300 ms;
Wired: 500 ms;
Wireless: 2000 ms
Wireless: 3000 ms
1 GB DDR3, 8 GB SDHC card
25 cm × 30 cm
(10" × 12") cassette
size
248 × 297.6
282 (w) × 332 (l)
× 15.5 (h)
1.8 (4.0)
Aluminum frame with carbon-fiber entrance window
OLED display with WLAN, LAN, battery, and sensor indicators
802.11n WLAN standard @ 5 GHz
Channel: WAP:
Stationary:
Stationary DFS:
GigE, trigger and power via optional docking connector
Integrated X-ray trigger control
Automatic Exposure Detection
65% (1 cy/mm), 35% (2 cy/mm), 10% (4 cy/mm) for RQA5
75% (0 cy/mm), 60% (1 cy/mm), 40% (3 cy/mm) for RQA5
www.vareximaging.com
Environmental Considerations
4336
Wired: 600 ms;
Wireless: 3600 ms
36 cm × 43 cm
(14" × 17") cassette
(17" × 17") cassette
size
350 × 426
384 (w) × 460 (l)
× 15.5 (h)
3.2 (7.0)
36-48, 149-165
36-48, 149-165
a
52-64, 100-116, 132-140
5.0 cy/mm
1%
4343
43 cm × 43 cm
size
426 × 426
460 (w) × 460 (l)
× 15.5 (h)
3.8 (8.4)
b
b
35
loading