HEIDENHAIN ROD 420 TTL Instrucciones De Montaje página 7

Ocultar thumbs Ver también para ROD 420 TTL:
Warnings
Fault detection signal:
Temperature at thermal switch > 100°C (± 5 K)
a)
¬ UaS = Low *
¬ Ausgangssignale weiterhin verfügbar
LED failure or contamination of the graduated disk
b)
¬ UaS = Low (min. 20 ms)
¬ Output signals at high impedance **
c)
d)
Attention: A temperature at the thermo switch of 110° C (± 5 K) switches off the supply voltage.**
* The fault-detection signal is not reset until after the rotary encoder has cooled.
** Operation with position feedback is then no longer possible.
U
aS
TTL
U
U
= High:
aS
aS
HTL
U
< 8 V
P
¬ UaS = Low
¬ Output signals at high impedance **
Temperature overload of cable driver
¬ UaS = Low *
¬ Output signals at high impedance **
= Low:
1 V
SS
7
loading

Este manual también es adecuado para:

Rod 426 ttlRod 430 htlRod 436 htlRod 480 1 vssRod 486 1 vss