TECHNICAL ALARM DISCRIPTION
The monitor will show 'HI' or 'Lo' as technical alarm on LCD with no delay if the
determined blood pressure (systolic or diastolic) is outside the rated range specified
in part SPECIFICACIONS. In this case, you should consult a physician or check if your
operation violated the instructions.
The technical alarm condition (outside the rated range) is preset in the factory and
cannot be adjusted or inactivated. This alarm condition is assigned as low priority
according to IEC 60601-1-8.
The technical alarm is non-latching and need no reset. The signal displayed on LCD will
disappear automatically after about 8 seconds.
-34-