Connect the patient cable to the reusable sensor:
1. Refer to Fig. 3. Orient the sensor connector to the patient cable connector.
2. Refer to Fig. 4. Insert the sensor connector completely into the patient cable connector. Close the sensor connector
cover over the patient cable connector until it locks in place.
C. Disconnect the patient cable from the sensors:
NOTE: To avoid damage, pull on the sensor connector, not the cable.
Disconnect the single-patient-use sensor:
1. Refer to Fig. 5. Pull firmly on the sensor connector to remove it from the patient cable.
Disconnect the reusable sensor:
1. Refer to Fig. 6. While holding the sides of the patient cable connector, lift the sensor connector cover from the patient
cable connector.
2. Refer to Fig. 7. Pull firmly on the sensor connector to remove it from the patient cable.
CLEANING
1. Remove the sensor from the patient and disconnect it from the patient cable.
2. Clean the surface of the cable by wiping it with a 70% isopropyl alcohol pad.
3. Wipe all surfaces of the cable.
4. Saturate another cloth or gauze pad with sterile or distilled water and wipe all surfaces of the cable.
5. Dry the cable by wiping all surfaces with a clean cloth or dry gauze pad.
CAUTION:
• Do not immerse the connectors on the cable in any liquid solution.
• Do not sterilize by irradiation, steam, autoclave, or ethylene oxide.
• Do not clean with chemicals not approved above.
ENVIRONMENTAL
Operating Temperature
Storage Temperature
Relative Humidity
COMPATIBILITY
RD rainbow SET series EMS patient cables are intended for use only with devices containing Masimo SET or Masimo
rainbow SET technology and with compatible sensors. Consult individual oximetry system manufacturers for
compatibility of particular devices and sensor models. Each device manufacturer is responsible for determining
whether their devices are compatible with each sensor and cable model. Use of this cable with other devices may
result in no or improper performance.
For Compatibility Information Reference: www.Masimo.com
WARRANTY
Masimo warrants to the initial buyer only that this product, when used in accordance with the directions provided with the
Products by Masimo, will be free of defects in materials and workmanship for a period of 24 months (2 years). Single use
products are warranted for single patient use only.
THE FOREGOING IS THE SOLE AND EXCLUSIVE WARRANTY APPLICABLE TO THE PRODUCTS SOLD BY MASIMO TO BUYER.
MASIMO EXPRESSLY DISCLAIMS ALL OTHER ORAL, EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION
ANY WARRANTIES OF MERCHANTABILITY OR FITNESS FOR PARTICULAR PURPOSE. MASIMO'S SOLE OBLIGATION AND
BUYER'S EXCLUSIVE REMEDY FOR BREACH OF ANY WARRANTY SHALL BE, AT MASIMO'S OPTION, TO REPAIR OR REPLACE
THE PRODUCT.
WARRANTY EXCLUSIONS
This warranty does not extend to any product that has been used in violation of the operating instructions supplied with the
product, or has been subject to misuse, neglect, accident or externally created damage. This warranty does not extend to
any product that has been connected to any unintended device or system, has been modified, or has been disassembled or
reassembled. This warranty does not extend to sensors or patient cables that have been reprocessed, reconditioned or recycled.
IN NO EVENT SHALL MASIMO BE LIABLE TO BUYER OR ANY OTHER PERSON FOR ANY INCIDENTAL, INDIRECT, SPECIAL OR
CONSEQUENTIAL DAMAGES (INCLUDING WITHOUT LIMITATION LOST PROFITS), EVEN IF ADVISED OF THE POSSIBILITY
THEREOF. IN NO EVENT SHALL MASIMO'S LIABILITY ARISING FROM ANY PRODUCTS SOLD TO BUYER (UNDER A CONTRACT,
WARRANTY, TORT OR OTHER CLAIM) EXCEED THE AMOUNT PAID BY BUYER FOR THE LOT OF PRODUCT(S) INVOLVED IN
SUCH CLAIM. IN NO EVENT SHALL MASIMO BE LIABLE FOR ANY DAMAGES ASSOCIATED A PRODUCT THAT HAS BEEN
REPROCESSED, RECONDITIONED OR RECYCLED. THE LIMITATIONS IN THIS SECTION SHALL NOT BE DEEMED TO PRECLUDE
ANY LIABILITY THAT, UNDER APPLICABLE PRODUCTS LIABILITY LAW, CANNOT LEGALLY BE PRECLUDED BY CONTRACT.
NO IMPLIED LICENSE
Purchase or possession of these patient cables does not carry any express or implied license to use these cables with any device
that is not an authorized device or separately authorized to use Masimo Patient Cables.
41°F to 104°F (5°C to 40°C)
-40°F to 158°F (-40°C to 70°C)
10% to 95% non-condensing