RD rainbow SET® Series
EMS Patient Cables
Reusable
INDICATIONS
The RD rainbow SET® EMS patient cables have the same indications for use as the applicable sensors. Please see the compatible
sensor's directions for use for indications for use and prescribed information.
DESCRIPTION
RD rainbow SET EMS patient cables are used with Masimo® RD rainbow™ and RD SET® sensors.RD rainbow SET EMS patient cables
are for use only with devices containing Masimo rainbow SET® technology Version 7.1 or higher or licensed to use rainbow®
compatible sensors. Consult individual oximetry system manufacturers for compatibility of particular sensor and device models.
Each device manufacturer is responsible for determining whether their devices are compatible with each sensor model.
RD rainbow SET EMS patient cables were verified using Masimo rainbow SET technology.
WARNING: Masimo sensors and cables are designed for use with devices containing Masimo SET® or Masimo rainbow SET
oximetry or licensed to use Masimo sensors.
WARNINGS, CAUTIONS, AND NOTES
• Always refer to the oximeter module operator's manual for complete instructions or additional instructions.
• Ensure the cable is physically intact, with no broken or frayed wires or damaged parts. Visually inspect the cable and discard
if cracks or discoloration are found.
• All sensors and cables are designed for use with specific monitors. Verify the compatibility of the monitor, cable and sensor
before use, otherwise degraded performance and/or patient injury can result.
• Carefully route patient cable to reduce the possibility of patient entanglement or strangulation.
• Failure to properly connect the sensor or the oximeter module to the cable will result in intermittent readings, inaccurate
results, or no reading.
• To avoid damage to the cable, always hold it by the connector rather than the cable when connecting or disconnecting
either end.
• To prevent damage, do not soak or immerse the sensor or cable in any liquid solution. Do not attempt to sterilize the cable
or sensor.
• Do not attempt to reprocess, recondition or recycle Masimo sensors or patient cables as these processes may damage the
electrical components, potentially leading to patient harm.
• Caution: Replace the cable when a replace cable message is displayed, or when a low SIQ message is consistently displayed
while monitoring consecutive patients after completing the low SIQ troubleshooting steps identified in the monitoring
device operator's manual.
• Note: The cable is provided with X-Cal® technology to minimize the risk of inaccurate readings and unanticipated loss of
patient monitoring. Replace the cable when the patient monitoring time is exhausted. Some legacy monitoring devices
do not distinguish between the cable and the sensor. In these devices, a replace sensor or similar message could pertain
either to the sensor or the cable.
CABLE AND SENSOR SELECTION
Note: The maximum combined length for the sensor, RD rainbow SET patient cable, and RD SET extension cable is 15 feet.
Refer to the table below for acceptable sensor and cable length combinations:
Cable Length
1.5 ft.
4 ft.
12 ft.
INSTRUCTIONS
A. Connect the patient cable to the device
1. Connect the device connector of the patient cable to the mating connector of the device.
2. Ensure the patient cable connector is fully engaged with the device. A tactile or audible click of connection should
be heard.
B. Connect the patient cable to the sensor
Connect the patient cable to the single-patient-use sensor:
1. Refer to Fig. 1. Orient the sensor connector to the patient cable connector as shown.
2. Refer to Fig. 2. Insert the sensor connector completely into the patient cable connector until it locks in place as shown.
D I R E C T I O N S FOR U S E
LATEX
Not made with natural rubber latex
PCX-2108A
02/13
0.5 ft.
YES
YES
YES
Sensor Length
2 ft.
YES
YES
YES
3
3 ft.
6 ft.
YES
YES
YES
YES
NO
YES
10825C-eIFU-0621
en
Non-Sterile