9. TROUBLESHOOTING

Problem
Any problem accompanied by infor-
mation on concentrator display, indi-
cator lights and/or audible signals
©2022 Inogen, Inc. All rights reserved.
Description
Battery HOT
The battery has exceeded tem-
perature limit while concentrator
is running on battery power. The
concentrator will shut down and
stop producing oxygen.
System HOT
Concentrator temperature is too
high. The concentrator will shut
down and stop producing oxygen.
Sensor fail
The concentrator's oxygen sensor
has malfunctioned.
System COLD
The system is cold (<2˚C). The con-
centrator will shut down and stop
producing oxygen.
System Error
The concentrator has stopped pro-
ducing oxygen and is shutting down.
Possible Cause
Refer to device icon & alarm
glossary
Page 32 of 356
What To Do
If possible, move concentrator to a
cooler location, then turn power off
and back on. Ensure air intake and
outlet vents have clear access and
particle filters are clean. If condition
persists, switch to external power
or a backup source of oxygen and
contact your equipment provider.
Ensure air intake and outlet vents
have clear access and particle
filters are clean. If condition per-
sists, switch to a backup source of
oxygen and contact your equipment
provider.
You may continue to use the con-
centrator. If the condition persists,
contact your equipment provider.
Move to a warmer environment to
allow the unit to warm up before
starting it. If condition persists,
switch to a backup source of oxy-
gen and contact your equipment
provider.
Switch to backup oxygen source
and contact your equipment
provider.
Recommended Solution
Refer to device icon & alarm
glossary