• If the skin tone identified by
the skin tone and contact
sensor is too dark, the flash
activation will be blocked,
and the status indicator will
turn "momentarily" red.
In case of general error:
• Reset the device by
unplugging the power cord
from the electrical outlet on
the wall, wait 10 seconds
and then reconnect it.
Then try turning on the
device after resetting.
• In case of dirt or debris on
the treatment window, you
should clean the treatment
window according to the
cleaning section. If you can't
get the area clean, contact
our customer service.
• In case of a pulse delay after
repositioning the device on a
different treatment area, you
should consider performing
the treatment in a cooler
environment. The device
may pulse at a slower rate
in a warm environment.
If the problem is not
solved, please contact
our customer service.
20

Customer Service

For more information about
the device, if your device is
broken, stopped working, or
if you need more assistance,
please contact us:
+44 0800-923-4833
Device specification
parameters
Manufactured by:
ELG trade Ltd.
Emitted light spectrum:
Large treatment window:
(yellow optical filter)
475nm to 1,200nm max.
Precise treatment window:
(red optical filter)
550nm to 1,200nm max.
Pulse duration:
5 ± 2 [ms]
Emitted energy flux:
5 [J/cm
]
2
Selectable energy levels:
6
Treatment area:
Large treatment window:
4.5 cm
2
Precise treatment window:
2 cm
2