5.10
Assembling the Handle
Position handle shells at an angle of about 45° to each other.
•
Engage the hooks provided on the upper side of the shells.
•
Now close the handle until the two shells touch each other.
•
Verify that the two shells are well aligned to each other, then press them together until
•
they click audibly in place.
5.11
Functional Check
CAUTION
Risk of injury from instruments or components that have not been checked!
Non-functional or defective instruments and components can endanger the patient or user,
as well as affect the intended function of the marSeal 5 with the currents
®
SealSafe
/SealSafe
Always keep electrosurgical instruments and their components in flawless working order.
•
Never use electrosurgical instruments and their components if they are damaged due to
•
improper handling or transport. If instruments or components are damaged, they must be
disposed of immediately.
Use of components or accessories by other manufacturers may constitute a source of
•
danger. In case of doubt, contact Gebrüder Martin.
76
Bipolar Sealing System marSeal 5 with sterile blade holder
®
IQ.
Push together tubular shaft (1)
•
and handle (7); if necessary,
press the two red unlock
pushbuttons (4).
Ensure that tubular shaft (1) and
•
handle (7) snap into place as
intended.
Then move the movable
•
handle (6) and press blade
trigger (5) to test the correct
assembly.
Instructions for Use
Revision 2