Assembling The Handle; Functional Check - KLS Martin group marSeal 5 Instrucciones De Uso

Sistema de sellado bipolar con soporte de cuchilla estéril
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 79
5.10

Assembling the Handle

Position handle shells at an angle of about 45° to each other.
Engage the hooks provided on the upper side of the shells.
Now close the handle until the two shells touch each other.
Verify that the two shells are well aligned to each other, then press them together until
they click audibly in place.
5.11

Functional Check

CAUTION
Risk of injury from instruments or components that have not been checked!
Non-functional or defective instruments and components can endanger the patient or user,
as well as affect the intended function of the marSeal 5 with the currents
®
SealSafe
/SealSafe
Always keep electrosurgical instruments and their components in flawless working order.
Never use electrosurgical instruments and their components if they are damaged due to
improper handling or transport. If instruments or components are damaged, they must be
disposed of immediately.
Use of components or accessories by other manufacturers may constitute a source of
danger. In case of doubt, contact Gebrüder Martin.
76
Bipolar Sealing System marSeal 5 with sterile blade holder
®
IQ.
Push together tubular shaft (1)
and handle (7); if necessary,
press the two red unlock
pushbuttons (4).
Ensure that tubular shaft (1) and
handle (7) snap into place as
intended.
Then move the movable
handle (6) and press blade
trigger (5) to test the correct
assembly.
Instructions for Use
Revision 2
Tabla de contenido
loading

Tabla de contenido