CAUTION
Risk of injury for patient and user from improper use!
Improper application as well as misuse can cause severe injuries to the patient and operator
and/or lead to premature wear of the sealing system!
Clean and sterilize marSeal 5 before each application, see section 5 "Reprocessing,
•
Cleaning, Disinfection and Sterilization", page 62.
A functional check must be carried out before each application, see section 5.11
•
"Functional Check", page 76!
When inserting the instrument into or withdrawing it from a trocar, the jaw part must be
•
closed!
Use the bipolar sealing system only for its intended purpose, see section 1.3 "Intended
•
Purpose", page 47.
There is danger of injury from the blade when inserting the blade holder!
•
•
Avoid contact with the blade!
•
Lay out the connecting cable in such a way that it touches neither the patient nor other
cables. Protect the cable against mechanical damage (squashing by trolley castors,
pinching, kinking, etc.)!
Use of high-frequency devices can interfere with other devices!
•
56
Bipolar Sealing System marSeal 5 with sterile blade holder
Instructions for Use
Revision 2