9. Error messages

The error messages are indicated at the rear of the control unit by the status LED (flashing green).
Flashing cycle Description of error
1x
Overheating
2x
Foot control
5x
Time-out (> 15 min)
6x
Handpiece
8x
System error
Solution
> Switch off conrol unit
> Allow the control unit to cool for at least 10 minutes
> Observe permissible ambient temperature/operating mode
> Release foot control
> Release foot control
(must not be active for longer than 15 minutes without interruption)
> Check tip
(full engagement, damage, torque)
> Dry the handpiece/supply hose
> Check plug-in connection of the handpiece/supply hose
> If the error message appears again, contact an authorised iM3 service
partner immediately.
> Start the device again
> Contact an authorised iM3 service partner.
Control Unit U8200, U8300