Safety Notes - iM3 ULTRA-S Instrucciones De Uso

Ocultar thumbs Ver también para ULTRA-S:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 42

5. Safety notes

> Before using the device for the first time, store it at room temperature for 24 hours.
> Check the device for damage and loose parts each time before using.
> Do not operate the device if it is damaged.
> Always ensure the correct operating conditions and cooling function.
> Always ensure that sufficient and adequate cooling is delivered and ensure adequate suction
(the exception are endodontics applications).
> In case of coolant supply failure, the device must be stopped immediately.
The exception to this is in endodontic applications, where coolant is not used.
Maximum operating time without coolant:
> Perform a test run each time before using.
> Never touch the patient and the electrical contacts on the device simultaneously.
> Check the parameter settings every time you restart.
> Make sure that the supply hose is dry. Moisture in the supply hose can lead to a malfunction (risk of short circuit).
> Do not look directly into the optic (LED) outlet.
> Replace damaged or leaking O-rings immediately.
> Do not twist, kink or squeeze the supply hose (risk of damage).
10
Control unit/Foot control/ Handpiece
> 2 minutes for the power range 1–30
> 30 seconds for the power range 31–40
Tabla de contenido
loading

Este manual también es adecuado para:

Ultra-stU8200U8300

Tabla de contenido