Operation
> Before every patient: Perform rinsing function for automatic internal cleaning of the coolant channels.
Ultra
Rinsing function for automatic internal cleaning of the coolant channels
24
Control unit U8200, U8300
Remove the handpiece from the supply
Set power between off and 0
Press foot control 3 times within 3 seconds
> Rinsing function active for 30 seconds
Cancelling the rinsing function with one of these options:
> Press foot control
> Adjust power regulator
Control Unit - Functions
U8200
U8300
✔
✔
Control Unit - Rinsing function