F1 Unt Configuration Of Secondary Units Of Measure (*); F2 Bl Display Backlighting; F3 Com Transmission Mode; F4 Pro Serial Commands - Scale House HLD Serie Manual De Instrucciones

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

7.2 F1 Unt CONFIGURATION OF SECONDARY UNITS OF MEASURE (*)

By pressing ZERO it's possible to choose the units of measure for the conversion function (see section 6.8), enabling
(on) or disabling (oFF) the suggested secondary units of measure:
1) The display proposes the first secondary unit of measure to be enabled or disabled.
2) Press TARE to change the options and confirm with ZERO, or press PRINT/M+ to exit without confirming and
pass to the following unit.
3) The display proposes the 2nd secondary unit to be enabled or disabled.
4) Proceed as described in point 2 up until the last proposed unit of measure.

7.3 F2 bL DISPLAY BACKLIGHTING

EL oFF backlighting always off
EL on
backlighting always on.
EL AU
backlighting on only when the weight is greater than 20d or when a key is pressed. When a key is pressed,
if the weight is less than 20d, the backlighting remains active for 5 seconds.

7.4 F3 CoM TRANSMISSION MODE

PRINTER PORT
One will be asked to select the port to be used, press TARE to select one of the suggested values:
S 232
RS232 port
S USB Not used in this application.
Select S 232 and press ZERO to confirm.
TRANSMISSION MODE
One will be asked to select the transmission mode, press TARE to select one of the suggested values:
P Prt
Manual transmission through the PRINT/M+ key.
P Cont continuous transmission with P Cont. protocol
P Auto Automatic transmission upon weight stability (when in totalisation mode).
WirELE Not used in this application.
Press ZERO to confirm.
BAUD RATE
One will be asked to select the baud rate, press TARE to select one of the suggested values: 600, 1200, 2400 4800,
9600. Press ZERO to confirm.
PRINTER TYPE
One will be asked to select the printer type, press TARE to select one of the suggested values:
tP
enables the printing with ASCII printer (for example TPR).
LP-50
Non used in this application.
Select tP and press ZERO to confirm.

7.5 F4 Pro SERIAL COMMANDS

The scale can be operated with serial commands to execute zero, tare and read weight, see section 8.2
By pressing ZERO it's possible to set an address for the scale, press TARE key to select one of the suggested values:
Normal no address, therefore the commands must be sent without address.
485
it will be required to set the address, from 1 to 99. Use MODE to select the value, or set a value using the %
key and the PCS key to move the active digit and then using the TARE key to increment a digit. Confirm
with ZERO key.

7.6 "tEcH" - SCALE CONFIGURATION

Press the TARE key to find the step "tEcH". By pressing ZERO it's possible to access the scale configuration menu.
One is asked to enter the password.
To enter the menu press the PCS, M+, and TARE keys, one after the other.