b) Tare cancellation: to eliminate the stored tare, press the [Tare/Zero] key. The tare weight is cancelled and the gross
weight is displayed. The tare light turns off.
III. CALIBRATION
(1) Enter in the calibration menu: press the [Clear/CAL] key for about 3 seconds: the instrument shows the A/D
converter points relative to the weight.
(2) press the [Clear/CAL] key, the display shows "F= xxx": select the maximum capacity, with [Set/Div] key;
(3) press the [Clear/CAL] key, the display shows "d=0.xx" : select the divisions of the scale, with [Set/Div] key;
(4) press the [Clear/CAL] key, the display shows "CAL-0" (zero calibration): clear the loading and wait for stable (the
arrow on "Stable" turned on).
(5) press the [Clear/CAL] key, the display shows "L-xx" (calibration with sample weight): press the [Set/Div] key for select
the weight calibration;
(6) Load the scale and wait the stable.
(7) Press the [Clear/CAL] key for save and exit.
Note: during the calibration, press the [Tare/Zero] key to return to the previous step or exit the calibration; if one
exits before the last step, the modifications will not be saved; only when one ends with step (7), the parameters are
updated and the modifications are saved.
IV. BACKLIGHT FUNCTION
1) Press the [Menu/Hold] key for about 3 seconds
2) Press the [Set/Div] key for set:
- "b-AUto" enable backlight when the weigh is instable;
- "b-off" disable backlight
3) Press the [Menu/Hold] key for save and exit.
V. SELECT UNIT OF MEASURE
1) Press the [Menu/Hold] key for about 3 seconds:
2) Press the [M+] key until of the display show "U - xx";
3) Press the [Set/Div] key for set:
- kg ("U - kg");
- Lb ("U - Lb").
4) Press the [Menu/Hold] key for save and exit.
NOTE: It will possible see the selection unit of measure on the right of the LCD display.
VI. AUTO SWITCH-OFF FUNCTION
1) Press the [Menu/Hold] key for about 3 seconds:
2) Press the [M+] key until of the display show "Off xx"
3) Press the [Set/Div] key for set:
- "Off- 0" for disable the function
4