39
APPENDIX
Guidance and manufacturer's declaration – electromagnetic emissions
The device is intended for use in the electromagnetic environments listed below, and should only be used
in such environments:
Emissions test
RF emissions
CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environments listed below, and should only be used
in such environments:
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Power frequency
magnetic field
IEC 61000-4-8
Compliance
Group 1
Class B
Not Applicable
Not Applicable
IEC 60601
test level
± 8 kV contact
discharge
± 15 kV air discharge
30 A/m
50 or 60 Hz
ENGLISH
Electromagnetic environment – guidance
RF energy is used only to maintain device's operation.
Therefore, its RF emissions are so low that it's not
likely to cause any interference in nearby electronic
equipment.
The device is suitable for use in all establishments,
including domestic establishments, and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Compliance level
± 8 kV contact
discharge
± 15 kV air discharge
30 A/m
50 or 60 Hz
Electromagnetic environment –
guidance
In the case of air discharge testing,
the climatic conditions shall be
within the following ranges:
Ambient Temperature:15°C~35°C
Relative Humidity: 30%~60%.
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical com-
mercial or hospital environment.