Frontier
TM
2000 Series Centrifuge
Hold the rotor with one hand and secure the rotor to the shaft by turning the fixing nut clockwise. Tighten the
fixing nut with enclosed rotor key (See figures 4-5)
We will provide a tool for nut rotor with centrifuge.
ATTENTION!
Check that the fixing screw is properly installed before each run. (See figure 5-6)
Do not operate the centrifuge with rotors or buckets which show any signs of corrosion or
mechanical damage.
Do not operate with extremely corrosive substances, which could damage the rotor,
buckets and materials.
In case of any questions, please contact the manufacturer!
3.1.2 Loading angle rotors
Rotors must be loaded symmetrically and with equal weight (See figure below). The adapter may only be
loaded with the appropriate vessels. The weight differences between the filled vessels should be kept as low
as possible. Therefore we recommend weighing them with a balance. This reduces the wear of the drive and
the acoustic operating noise.
The maximum load per hole is stated on each rotor.
Figure.7 WRONG
Figure.8 CORRECT (6 tubes)
EN-5