Troubleshooting - Dr. Brown's Customflow BF100 Manual Del Usuario

Extractor de leche doble eléctrico
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Troubleshooting

ISSUE
POSSIBLE CAUSES
Power Adapter
No
Power
No
Tubing
Suction
Tubing
Assembly error
Low
Suction
Duckbill valve
Softshape™
Silicone Shield
Silicone
membrane
Too high suction
Pain
CHECKS
Check if the power cord is not
connected or it is damaged.
Check that wall outlet is
working properly.
Check connection to milk
collection kit and to pump port(s)
on the front of the unit.
Inspect tubing for holes or cracks.
Check for loose assembly of milk
collection kit components.
Check for damage or wear from use.
Check for damage.
Softshape™ Silicone Shield is
completely on the breast.
Check for damage or wear from use.
Check the vacuum level.
SOLUTION
Con rm connection of
power cord; replace damaged
power cord, move to a
working wall outlet.
Con rm connections.
Replace tubing if crack is in
the middle of the length.
If cracked near tube ends,
remove end and trim away
damaged tube
and reinstall tube end.
Disassemble and reassemble.
Take care to tighten the
membrane cap.
Replace valve if damaged or worn.
It is recommended to replace
valves every 1-3 months
depending on frequency of use.
Replace if damaged.
Press more rmly against
the breast.
Replace membrane if damaged or
worn. It is recommended to replace
membranes every 3-6 months
depending on frequency of use.
Set to the lowest suction level or
consult your lactation consultant
or medical professional.