Electromagnetic Emissions; Guidance And Manufacturer's Declaration - Electromagnetic Immunity - Dr. Brown's Customflow BF100 Manual Del Usuario

Extractor de leche doble eléctrico
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Electromagnetic Emissions

The Dr. Brown's Natural Flow® Custom ow™ Double Electric Breast Pump is intended for use in the electromagnetic
environment speci ed below. The user of the Breast Pump should assure that it is used in such an environment.
Guidance and manufacturer's declaration – electromagnetic emissions
Emissions
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage uctuations/
icker emissions
IEC 61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines IEC 61000-4-11
Power frequency (50/60 Hz)
magnetic eld IEC 61000-4-8
Compliance
The Breast Pump uses pull of vacuum only for its internal function. Therefore,
Group 1
its RF emissions are very low and are not likely to cause any interference in
Class B
Class A
The Breast Pump is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage power
Not Applicable
IEC 60601
test level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for
input/output lines
± 1 kV line(s) and neutral
<5 % UT (>95 % dip in UT)
for 0,5 cycle
40 % UT (60 % dip in UT)
for 5 cycles
70 % UT (30 % dip in UT)
for 25 cycles
<5 % UT (>95 % dip in UT)
for 5s
3 A/m
NOTE: UT is the AC mains voltage prior to application of the test level
Electromagnetic environment – guidance
nearby electronic equipment.
supply network that supplies buildings used for domestic purposes.
Compliance level
±6 kV contact
±8 kV air
Not Applicable
Not Applicable
Not Applicable
3 A/m
Electromagnetic
environment – guidance
Floors should be wood, concrete or
ceramic tile. If oors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Mains power quality should be that of a
typical commercial or hospital
environment
Mains power quality should be that of a
typical commercial or hospital
environment
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the Breast
Pump requires continued operation
during power mains interruptions, it is
recommended that the Breast Pump
be powered from an uninterruptible
power supply or a battery.
Power frequency magnetic elds should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
22