Long press on pushbutton > 6 s turn off device.
Self-test of device can be activated with shorting of tips (probes) while device is
turned off.
Device indicates continuity by LED "Rx" 6 and beeps if resistance between
tips is less than 500 kΩ for < 500 kΩ range, and if resistance is less than 10 Ω
for < 10 Ω range.
Continuity tester is intended for use on non-energized environment. The ELV
LED is used to signalize dangerous voltage. If ELV LED turns on during test,
testing must be stopped.
If low battery LED turns on 3 , 2x AAA batteries needs to be replaced.

Maintenance

When using the instrument in compliance with the instruction manual, no special
maintenance is required.

Cleaning

If the instrument is dirty after daily usage, it is advised to clean it by using a humid cloth
and a mild household detergent. Prior to cleaning, ensure that instrument is discon-
nected from external voltage supply and any other instruments connected. Never use
acid detergents or resolvent for cleaning.
EN
19