SciCan STATIM 2000S Manual Del Operador página 36

Ocultar thumbs Ver también para STATIM 2000S:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

7. Troubleshooting con't.
Cycle interrupted — NOT STERILE,
Cycle aborted — NOT STERILE and
CYCLE FAULT messages.
If kinked, straighten the tube. If the tube cannot
be straightened, remove it from the push-in fitting
attached to the Statim. Depress the collar on the
fitting and with the other hand pull firmly on the tube.
Once the tube is free of the fitting, cut the damaged
section of tubing away using a sharp instrument. Be
sure that you leave enough tube to reach the unit
when you re-attach the tube to the exhaust fitting. If
the tube is too short to remove a section, contact your
SciCan dealer fora replacement.
Make sure the compressor is working. To check,
remove the exhaust tube from the waste bottle.
Start the Air Drying Only Cycle, and place the free
end into a glass of water. If there is not a strong,
steady flow of bubbles, the compressor is not
functioning properly. Contact your SciCan dealer.
Wait a few minutes and attempt another cycle before
proceeding to the next solution. Remove the cassette.
Be careful. The metal parts will be hot and the
cassette will contain hot steam. Inspect the cassette
to ensure that the holes in the back of the seal are
perfectly aligned, and that the flexible lip of the seal is
completely free. Check the exhaust tube for kinks or
obstructions. If kinked, straighten the tube. If the tube
cannot be straightened, remove it from the push-in
fitting attached to the Statim. Depress the collar on the
fitting and, with the other hand pull firmly on the tube.
Once the tube is free of the fitting, cut the damaged
section of tubing away using a sharp instrument. Be
sure that you leave enough tube to reach the unit
when you re-attach the tube to the exhaust fitting. If
the tube is too short to remove a section, contact your
SciCan dealer for a replacement.
Check that the Statim has not inadvertently been
exposed to any electrical interference.
Refer to the Installation section dealing with
Environmental Considerations. (Section 3.1)
Try running another cycle. If the problem persists,
record the cycle fault message number and contact
your dealer.
Page 36
36

Capítulos

Tabla de contenido
loading

Productos relacionados para SciCan STATIM 2000S

Este manual también es adecuado para:

Statim 5000s

Tabla de contenido