Battery Operation; Technical Data; Indicator - KERN KFB-TM Manual De Instrucciones E Instalación

Ocultar thumbs Ver también para KFB-TM:
Tabla de contenido
2.2.8
Real time clock
If it is available in the instrument, the real time clock can be activated to get printout with day and
time information.
2.2.9
Operator information messages
The weight indicator has a number of general and diagnostic messages which are described in detail in
the user's guide.
2.2.10
Software version
The software revision level is displayed during the power-up sequence of the instrument.
The approved software version is 1.07 and 1.08.
The software version 1.08 includes possibility of multi-range.
2.2.11
Totalisation
The indicator can be configured with a totalisation function, adding actual weight display values to the
memory when pressing "M+" key if the equilibrium is stable.
Pressing "MR" key displays the total accumulated weight.
Pressing "M+" and "MR" key will clear the totalised value.
2.2.12

Battery operation

The indicator can be operated from an internal rechargeable battery, if this option is installed.
3.

Technical data

The KFN-TM / KFB-TM / BFB / IFB / SFB / UFB / UFN / NFB / BFN / NFN weighing instruments
are composed of separate modules, which are set out as follows:
3.1

Indicator

The indicators have the following characteristics:
Type:
Accuracy class:
Weighing range:
Maximum number of Verification
Scale Intervals:
Maximum tare effect:
Fractional factor:
Minimum input voltage per VSI:
Excitation voltage:
Circuit for remote sense:
Minimum input impedance:
Maximum input impedance:
Mains power supply:
Operational temperature:
Issued by DELTA
Certificate of EU type-approval no. DK 0199.202 rev.1
KFN-TM / KFB-TM / BFB / IFB / SFB / UFB / UFN / NFB
/ BFN / NFN
III and IIII
Single-interval, multi-range (2 ranges) or multi-interval (2
partial intervals)
≤ 6000 (class III), ≤ 1000 (class IIII) for single-interval
≤ 3000 (class III), ≤ 1000 (class IIII) for multi-range and
multi-interval
-Max within display limits
p'i = 0.5
1 µV
5 VDC
present on the model with 7-terminal connector
87 ohm
1600 ohm
9 - 12 VDC / 230 VAC, 50 Hz using external adapter
-10 °C to +40 °C
Annex page 4 of 14