7.
Specifications
Measurement
range
Conductivity
Temperature
Resolution
Conductivity
Temperature
Limits of error
Conductivity
Temperature
Power require-
ments
Size/weight
Display
Signal input
IP rating
Battery life
Ambient conditions
METTLER TOLEDO Education Line conductivity meter
Five_cond_edu.indb 13
All manuals and user guides at all-guides.com
Education Line conductivity meter
FE30
0.0 μS/cm...99.9 mS/cm
0° C ... 00 °C
Auto range
0.0 μS/cm ... 9.99 μS/cm
20.0 μS/cm ... 99.9 μS/cm
200 μS/cm ... 999 μS/cm
2.00 mS/cm ... 9.99 mS/cm
20.0 mS/cm ... 99.9 mS/cm
0.°C
± 0.5 % of measured value
± 0.3 °C
9 V DC, 0.4 W ± 0%
For use with CSA Certified (or equivalent appro-
ved) power source, which must have a limited
circuit output.
200 x 75 x 52 mm / 0.6 kg
Liquid crystal
Mini-Din
For indoor use only.
-
Temperature:
5...40 °C
Relative humidity: 5 %...80 % (non-
condensing)
Height up to 2000 m
Installation category: II
Pollution degree: 2
Specifications
Education Line conductivity meter
FG3
0.0 μS/cm...99.9 mS/cm
0° C ... 00 °C
Auto range
0.0 μS/cm ... 9.99 μS/cm
20.0 μS/cm ... 99.9 μS/cm
200 μS/cm ... 999 μS/cm
2.00 mS/cm ... 9.99 mS/cm
20.0 mS/cm ... 99.9 mS/cm
0.°C
± 0.5 % of measured value
± 0.3 °C
Rating:
6VDC,5mA
Powered with batteries: 4xAAA/LR3
.5V or NiMH .2 rechargeable
69 x 82 x 36 mm / 0.8 kg
(without batteries)
Liquid crystal
Mini-Din
IP54 for system
>250 working hours
Temperature:
5...40 °C
Relative humidity: 5 %...80 %
(non-condensing)
Height up to 2000 m
Installation category: II
Pollution degree: 2
3
13.03.2007 16:28:05
Tabla de contenido
loading

Este manual también es adecuado para:

Education line el3

Tabla de contenido