Type
Measurement method
Measurement range
Measurement accuracy
Power supply
Number of
measurements
Classification
Clinical test
EMC
Memory
Operating conditions
Storage condition
Dimensions
Weight
Applied part
Useful life

Technical Data

UA-767F/FAC
Oscillometric measurement
Pressure: 0 - 299 mmHg
Systolic pressure: 60 - 279 mmHg
Diastolic pressure: 40 - 200 mmHg
Pulse:
40 - 180 beats / minute
Pressure: ±3 mmHg
Pulse:
±5%
4 x 1.5V batteries (R6P, LR6 or AA) or
AC adapter (TB-233)
UA-767F includes 4 batteries.
UA-767FAC includes an AC adapter.
Approx. 700 times
Approx. 200 times
With pressure value 180 mmHg, room temperature
73.4 °F (23 °C.)
Internally powered ME equipment (Supplied by
batteries) / Class II (Supplied by adapter)
Continuous operation mode
According to ANSI / AAMI SP-10 1992
IEC 60601-1-2: 2007
Last 60 measurements for each of the four user
+50°F to +104°F (10°C to 40°C) / 15 to 85 %RH /
800 to 1060 hPa
-4°F to +140°F (-20°C to 60°C) / 10%RH to
95%RH
Approx. 5.5" (140mm) [W] x 2.3" (60mm) [H]
x 4.1" (105mm) [D]
Approx. 9.0 oz. (255 g), excluding the batteries
Cuff:
Type BF
Device:
5 years (when used six times a day)
Cuff:
2 years (when used six times a day)
English 23
LR6 (alkaline batteries)
R6P (manganese batteries)