No.
Type of
Causes of error
error
Open
frontal
Frontal contact is damaged
Alarm
8
"
WINDOW
" (see
OPEN
Chapter 4.5
Main control is damaged
pto. D)
Light not working properly
Light does
9
Fluorescent lights not
not turn on
working
Ballast is damaged
Glass not in closed position
UV
germicidal
lamp does
10
UV lamp does not work
not turn on
(optional
accessory)
Ballast is damaged
Front position security
micro activated
Intermittent
11
alarm
activated
Main control is damaged
Power cord cable
disconnected
Display
12
does not
No power
turn on
Main control is damaged
WARNING
All repair and maintenance work in the cabinet should be done by qualified technical staff
with knowledge about the maintenance procedures required by the cabinet.
UM-BIOIIADVANCE-EN-1302
All manuals and user guides at all-guides.com
Correction of error
Slide down the glass to closed position and
check that the frontal contact is not loose
and remove any type of obstruction.
Notify the Technical Service to receive
instructions for adjusting the contact.
Call a specialised technician to replace the
control.
Check that the connection in the control
card is correct (Annex A7).
Slightly turn the fluorescent light to ensure
good contact.
Change the fluorescent bulb.
Change the ballast.
Check the front slides properly.
Check the lamp holder is connected
properly.
Check that the connection in the control
card is correct (Annex A7).
Change the UV lamp.
Change the ballast.
Open the front and check the security
micro is not loose.
Call the Technical Service to receive
instructions on adjusting micros.
Call a specialised technician to replace the
control.
Check that power cord cable is properly
connected in cabinet and in socket on the
wall.
Check voltage on the socket of the wall.
Call a specialised technician to replace the
control.
33