For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture's declaration - electromagnetic immunity
The 04-695-001 is intended for use in the electromagnetic environment
specified below. The customer or the user of the 04-695-001 should assure
that it is used in such an environment.
IMMUNITY test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Table 2
IEC 60601
test level
± 6 kV contact
± 8 kV air
3 A/m
23
Compliance level
± 6 kV contact
± 8 kV air
3 A/m
Electromagnetic
environment
guidance.
Floors should be
wood, concrete
or ceramic tile. If
floors are covered
with synthetic
material, the
relative humidity
should be at least
30%.
Power frequency
magnetic
fields should
be at levels
characteristic of
a typical location
in a typical
commercial
or hospital
environment.