This device complies with Part 15 of the FCC Rules. Operation is subject to the
following two conditions: (1) This device may not cause harmful interference, and (2)
this device must accept any interference received, including interference that may
cause undesired operation.
ELECTROMAGNETIC COMPATIBILITY INFORMATION
Guidance and manufacture's declaration - electromagnetic emission
The 04-695-001 is intended for use in the electromagnetic environment specified
below. The customer or the user of the 04-695-001 should assure that it is used in
such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Table 1
For all ME EQUIPMENT and ME SYSTEMS
Compliance
Group 1
Class B
Not Applicable
Electromagnetic
environment guidance.
The 04-695-001 uses RF
energy only for its internal
function. Therefore, its
RF emissions are very
low and are not likely to
cause any interference
in nearby electronic
equipment.
The 04-695-001 is
suitable for use in all
establishments other
than domestic and those
directly connected to the
public low-voltage power
supply network that
supplies buildings used
for domestic purposes.