Guidance And Manufacturer's Declaration - DeVilbiss DV57 Sleepcube AutoBilevel CPAP Serie Guía De Instrucciones

Dispositivo de presión positiva en las vías respiratorias
Ocultar thumbs Ver también para DV57 Sleepcube AutoBilevel CPAP Serie:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 23
Additional Specifications
Equipment classification with respect to protection from electric shock ........................................................................ Class II
Degree of protection from electric shock ................................................................................................. Type BF Applied Part
Degree of protection against ingress of liquids ......................................................................................IPXI Drip-proof vertical
Mode of operation .................................................................................................................................................... Continuous
Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide
DEViLbiSS GuiDANcE AND MANuFActuRER'S DEcLARAtioN
WARNiNG
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into
service according to the Electromagnetic Compatibility [EMC] information provided in the accompanying
documents.
Portable and Mobile RF Communications Equipment can affect Medical Electrical Equipment.
The equipment or system should not be used adjacent to or stacked with other equipment and that if adjacent or
stacked use is necessary, the equipment or system should be observed to verify normal operation in the
configuration in which it will be used.
NOTE– The EMC tables and other guidelines provide information to the customer or user that is essential in determining
the suitability of the Equipment or System for the Electromagnetic Environment of use, and in managing the
Electromagnetic Environment of use to permit the Equipment or System to perform its intended use without disturbing
other Equipment and Systems or non-medical electrical equipment.
Guidance and Manufacturer's Declaration – Emissions All Equipment and Systems
This device is intended for use in the electromagnetic environment specified below. The customer or user of this device
should assure that it is used in such an environment.
Emissions Test
Compliance
RF Emissions
Group 1
CISPR 11
Class B
RF Emissions
Radiated and Conducted
CISPR 11
Emissions
Harmonics
Class A
IEC 61000-3-2
Flicker
Complies
IEC 61000-3-3
Immunity Test
IEC 60601 Test Level
Electrostatic
±6kV contact
Discharge (ESD)
±8kV air
IEC 61000-4-2
Electrical Fast
Transient/burst
±2kV on AC Mains
IEC 61000-4-4
Surge
±1kV Differential
IEC 61000-4-5
±2kV Common
>95% Dip for
0.5 Cycle
Voltage dips, short
60% Dip for
interruptions and
5 Cycles
voltage variations
on power supply
30% Dip for
input lines
25 Cycles
IEC 61000-4-11
>95% Dip for
5 Seconds
SE-DV57-1
Electromagnetic Enforcement – Guidance
This device uses RF energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to cause any interference
in nearby electronic equipment.
This device is suitable for use in all establishments including domestic,
and those directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes.
Compliance Level
Electromagnetic Environment - Guidance
Floors should be wood, concrete or ceramic
±6kV contact
tile. If floors are synthetic, the relative humidity
±8kV air
should be at least 30%.
Mains power quality should be that of a typical
±2kV on AC Mains
commercial or hospital environment.
±1kV Differential
Mains power quality should be that of a typical
±2kV Common
commercial or hospital environment.
>95% Dip for
Mains power quality should be that of a typical
0.5 Cycle
commercial or hospital environment. If the user
60% Dip for
of this device requires continued operation
5 Cycles
during power mains interruptions, it is
30% Dip for
recommended that the device be powered
25 Cycles
from an uninterruptible power supply or
>95% Dip for
battery.
5 Seconds
EN - 15
Tabla de contenido

Solución de problemas

loading