Specifications - DeVilbiss DV57 Sleepcube AutoBilevel CPAP Serie Guía De Instrucciones

Dispositivo de presión positiva en las vías respiratorias
Ocultar thumbs Ver también para DV57 Sleepcube AutoBilevel CPAP Serie:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 23

SpEciFicAtioNS

cpAp
Size ........................................................................................................... 4.2"H x 6.5"W x 6.9"D (10.7cm x 16.5cm x 17.5cm)
Weight ............................................................................................................................................................. 1.22 kg (2.7 lbs.)
Electrical Requirements AC ...................................................................................................................... 100-240V~, 50/60 Hz
Electrical Requirements DC .................................................................................................................. 10.4 - 15 VDC, 5 Amps
Maximum Power Consumption ......................................................65 watts max from AC power source (flow generator only)
Bilevel Pressure Range ...........................................................................................................................................3-25 cmH
Operating Temperature Range ....................................................................................................... 41˚F to 104˚F (5˚C to 40˚C)
Operating Humidity Range .......................................................................................................0% to 95% RH non-condensing
Operating Atmospheric Conditions ..........................................................................................Sea level to 8,500 feet (2600 m)
Storage & Transportation Temperature Range ........................................................................-40˚F to +158˚F (-40˚C to 70˚C)
Storage & Transportation Humidity Range ...............................................................................0% to 95% RH non-condensing
Maximum Limited Pressure ...........................................................................................................30 cmH
Sound Pressure Level (tested per ISO 17510) ...............................................................................................................26 dBA
Sound Power Level .........................................................................................................................................................38 dBA
SmartCode
Technology is protected under US patent 8649510
®
Humidifier
Humidity Output (in the operating flow range): ................................................................................................ ≥10 mgH
Size: ......................................................................................................2.6" H x 6.3" W x 8.4" D (6.6 cm x 16.0 cm x 21.3 cm)
Weight: ........................................................................................................................................................1.75 lbs. (0.794 Kg)
Power Consumption: .................................................................................................................................................... 85 Watts
DV5HH Electrical Requirements ................................................................................................................100-240VAC, 0.95 A
Heater Plate Thermal Cutout:.............................................................................................................................. 305˚F (152˚C)
Power/Temperature Control: ......................................................................................Setting of 1 (minimum) to 10 (maximum)
Heater Plate Temperature: .................................................................................84˚F to 149˚F (approx. 29˚C to approx. 65˚C)
Operating Temperature Range: ...................................................................................................... 41˚F to 104˚F (5˚C to 40˚C)
Operating Humidity Range: .......................................................................................................0 to 95% R.H. non-condensing
Operating Atmospheric Pressure Range: ........................................................................................................70.0 – 106.0 kPA
Filter Specifications
Standard Filter .........................................................................................................................................> 3.0 micron particles
Optional Fine Particle Filter .....................................................................................................................> 0.3 micron particles
Dynamic Short-term pressure Accuracy with and without DV5HH Humidifier (per
iSo 17510-1:2007)
Waveform
Volume [mL]
Sinusoidal
cycle with I:E =
1:1
Maximum Flow-Rate (per iSo 17510-1:2007)
Measured pressure at the patient
connection port (cmH
O)
2
Average flow at the patient connection port
(l/min)
Long Term Static Pressure Accuracy ................................................................................................................. +/- 0.5 cmH
EN - 14
Breath Rate [min-1]
500
10, 15, 20
3.0 cmH
Set Pressure
[cmH
3.0
9.0
14.0
20.0
25.0
Test Pressures
O
9.0 cmH
O
2
2
3.0
8.9
95.0
160.5
Pressure Accuracy
O]
[Pk-Pk cmH
2
14.0 cmH
O
20.0 cmH
2
13.0
19.0
172.2
158.5
O
2
O under normal use
2
O/l air
2
O]
2
0.5
0.5
1.0
1.0
1.0
O
25.0 cmH
O
2
2
24.0
139.7
O
2
SE-DV57-1
Tabla de contenido

Solución de problemas

loading

Tabla de contenido