Agilent Technologies Turbo-V 550 Serie Manual De Instrucciones página 262

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 55
17
Technical Information
Optional Serial Port
When the Request is "J" (Numerical readings), the Answer will
contain a string of 5 characters with the following parameters:
Tab. 10
BYTES
1
2
3
4
5
When the Request is "K" (Counters readings), the Answer will
contain a string of 11 characters with the following parameters:
Tab. 11
BYTES
1-4
5-8
9-10
11
When the Request is "L" (software version), the Answer will contain
a string of 12 characters with the following parameters:
Tab. 12
BYTES
1-3
4-7
8-11
12
On request a sample program in QBasic language is available by
Agilent.
262/310
MEANING
Current (0-255 scaled)
Voltage (0-255 scaled)
Rotational speed KRPM
Pump temperature °C
(0-70 temperature reading, 255 = fail)
CRC
MEANING
Cycle time
Pump life
Cycle number
CRC
MEANING
Software version
CRC 16 EPROM
CRC 16 parameters
CRC
Turbo-V 550 Controller User Manual / 87-900-873-01 (G)
Tabla de contenido
loading

Este manual también es adecuado para:

969-9444969-9544