Sub menu "Service"
▲
▼
This menu is for maintenance authorised personnel only. The parameters stored in this menu can only be
displayed but not modified... with the exception of the language.
Sterilisation cycle
1. To start a cycle, select the desired sterilisation programme
2. Press the START button
3. The display will show on the upper row the selected cycle (temperature and time), while on the lower row
will show the actual product temperature and the remaining sterilisation time. In order to see the vessel
temperature, press MENU and use the arrows in order to see both Vessel and Product temperature.
P
4. Once the sterilisation temperature has been reached, the sterilisation time count will begin. At the end of the
sterilisation time a buzzer is activated
5. In the cooling process, when the vessel temperature goes below 100 °C the LED END CYCLE will light up
6. Once the cooling process is completed, open the lid with the handle
7. Switch off the autoclave via the main switch (Pos. 1 page 7)
NOTE:
Press the STOP button to end a running cycle.
WARNING! STOP button is an emergency button. Pushing it will abort the cycle and will open the
valves. We don't recommend you push it unless a problem occurs during the cycle.
The same thing will happen if the instrument is turned off by the main switch.
The set values are stored in the memory will not be lost after switching the instrument off.
Validation
The autoclave should be checked each time with chemical indicator and validated monthly with biological
indicators.
S
E R V
I
C E
1
1
2
1
C
1
0
3
.
4
15
-
1
0 m
C
-
1
0 m