VWR Vapour-Line Eco 50 Manual De Usario página 12

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 88
If the printer is enabled the "PRINTER ON" LED on the control panel will be lit.
Printing intervals can be set from a minimum of 1 minute up to a maximum of 30 minutes with a 1 minute step.
Sub menu "F0"
F0 parameter settings can be modified in this sub menu. The mathematical expression used for the F0
calculation (in minutes) is as follows:
where T is the vessel temperature, z is a microrganism-related parameter ranging from 4,0 to 15,0
(default 10), ∆t is the time interval between measurements. The calculation of F0 starts after a threshold
temperature Ts, ranging from 100 to 120 °C (default 100 °C), is reached.
From the F0 menu, press ENTER.
12
Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

481-0665

Tabla de contenido