R&S
FPC1000
Instrument damage caused by cleaning agents
Cleaning agents contain substances that may damage the instrument. For
example, cleaning agents that contain a solvent may damage the front
panel labeling, plastic parts, or the display.
Never use cleaning agents such as solvents (thinners, acetone, etc), acids,
bases, or other substances.
The outside of the instrument can be cleaned sufficiently using a soft, lint-
free dust cloth.
RF input
The RF input with an impedance of 50 Ω allows you to connect a DUT to the
R&S FPC1000. Typically, you connect the DUT with a cable and an appropriate
connector (for example a male N connector).
The frequency range of the RF input is specified in the datasheet.
The attenuation range is between 0 dB and 40 dB.
Risk of instrument damage
Make sure not to overload the RF input and keep within the maximum
allowed signal levels. Refer to the datasheet for the maximum allowed sig-
nal levels.
A DC input voltage of 50 V must never be exceeded.
Power switch
The power switch turns the R&S FPC1000 on and off when it is supplied with
power.
For more information, see
on page 9.
Headphone jack
The female headphone jack allows you to connect headphones (or external
speakers) with a miniature jack plug.
Getting Started 1178.4124.02 ─ 03
Chapter 2.4, "Turning the R&S FPC1000 on and
Instrument Tour
Front Panel
off",
12