Slope Calibration - ProMinent DULCOTEST ICT 1-IMA-1 m Instrucciones De Servicio

Medidor de conductividad inductivo
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

ProMinent
®
• Calibrate the zero point only when the sensor tip is completely
dry.
• Calibrate the zero point in the air when dismantled only.
• Hold the sensor tip more than 20 mm away from all objects
during calibration.
• Calibrate the zero point before the slope.
To calibrate the zero point of the sensor, dismantle the sensor
and then completely dry the sensor tip and the sensor shaft.
Select the first menu option in the "cali. zero point" in the setting
menu and wait approx. 30 s until the measured value is stable –
then press Enter.
8.3

Slope calibration

IMPORTANT
• Always follow the handling instructions below.
• During the calibration the water sample must be flowing
evenly without bubbles or turbulence or at a standstill.
• If calibrating in flowing water, the electrolytic conductivity of
the water sample must be constant.
• The axis through the bore of the sensor tip must line up with
the direction of the flow.
Cover the sensor tip completely with the water sample (immerse
or flood the environment)
There are now three alternative procedures for different require-
ments:
Commissioning
35
Tabla de contenido
loading

Este manual también es adecuado para:

Dulcotest ict 1-ima-2 m

Tabla de contenido