Garantía y declaración de conformidad
8.3 Directrices CEM y declaración del fabricante
Guidance and manufacturer's declaration - electromagnetic emissions
The LM-ProPower is intended for use in the electromagnetic environment specified below.
The customer or the user of the LM-ProPower should assure that it is used in such an environment.
Emission test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Guidance and manufacturer's declaration - electromagnetic immunity
The LM-ProPower is intended for use in the electromagnetic environment specified below.
The customer or the user of the LM-ProPower should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage varaiations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE
U
is the a.c. mains voltage prior to application of the test level.
T
30
Compliance
Group 1
Class B
Not applicable
Not applicable
IEC 60601
test level
±6 kV contact
±6 kV contact
±8 kV air
±8 kV air
±2 kV for power
±2 kV for power
supply lines
supply lines
±1 kV for input/output
lines
lines
±1 kV differential
±1 kV differential
mode
mode
±2 kV common mode
±2 kV common mode
<5 % U
<5 % U
T
(>95 % dip in U
)
(>95 % dip in U
T
for 0,5 cycle
for 0,5 cycle
40 % U
40 % U
T
(60 % dip in U
)
(60 % dip in U
T
for 5 cycles
for 5 cycles
70 % U
70 % U
T
(30 % dip in U
)
(30 % dip in U
T
for 25 cycles
for 25 cycles
<5 % U
<5 % U
T
(>95 % dip in U
)
(>95 % dip in U
T
for 5 sec
for 5 sec
3 A/m
3 A/m
Electromagnetic environment - guidance
The LM-ProPower uses RF energy only for its internal function.
Therefore, it's RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
The LM-ProPower is suitable for use in all establishments, including
domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domestic purposes.
Compliance level
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a
typical commercial or hospital
environment.
±1 kV for input/output
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
T
typical commercial or hospital
)
T
environment. If the user of the
LM-ProPower requires continued
operation during power mains i
T
interruption, it is recommended that the
)
T
LM-ProPower be powered from an
uninterruptible power supply or battery.
T
)
T
T
)
T
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
Electromagnetic environment -
guidance