5.11.3.14
5.11.3.15
5.11.3.16
5.11.3.17
774 Oven Sample Processor, Instrucciones para el uso
5.11 Description of the remote control commands
Info.ActualInfo.Inputs.Status
Status of the Input lines (Input0...7) of the Remote interface.
$Q sends the signal state as decimal number e.g.
10 ⇒ 00001010 binary ⇒ 2
(active = low, inactive = high)
See also page 119ff.
Info.ActualInfo.Outputs.Status
Status of the output lines (Output0...13) of Remote interface. See 5.11.3.14.
Info.ActualInfo.Display.L1
Info.ActualInfo.Display.L2
Text of the first and second line of the LCD display.
Info.ActualInfo.Counter.Sample
Info.ActualInfo.Counter.Maximum
Currently processed sample number and total number of samples.
1
3
⇒ Input1 and Input3 active
+ 2
read only
read only
read only
read only
read only
read only
121