5 Descripción detallada
5.11.2.11
5.11.2.12
5.11.2.13
118
Configuration of the dosing units. Depending on the selection of the dosing
unit (see WetPartNo 5.11.2.9) the corresponding data set is overlayed.
These settings are only relevant for Dosinos. For each port of the Dosino
individual settings are possible.
MaxRate:
Max. possible dosing and filling speed allowed
Length:
Tubing length on selected port
Diameter:
Inner tubing diameter on selected port
Config.RSset
$G effects all the RS settings. Modifications are only possible in inactive in-
strument state. After setting the interface parameters wait for 2 sec. to allow
the components to equilibrate.
Config.RSset.Baud
Config.RSset.DataBit
Config.RSset.StopBit
Config.RSset.Parity
Config.RSset.Handsh
Settings for data transmission via RS interface, baud rate, data bit, stop bit,
parity and type of handshake, see also page 129ff.
Config.RSset.CharSet
Setting the character set and print mode. For data communication with
computers select 'IBM' (IBM code page 437).
300, 600, 1200, 2400, 4800, 9600
even, odd, none
HWs, HWfull, SWchar, SWline, none
IBM, HP, Epson, Seiko, Citizen